Recent
improvements of the Kloften & Kloften
Strip Splicer
January 16,
1996
Issue No.8
1. Two Step Splicing.
The method
described in Issue No. 7 has been
improved further by introducing an
air cylinder for the movement of the upper electrode holder from the
rear to the
front position.
The procedure
is now as follows:
o When the power is turned on, the
holder goes to the rear position.
o After completion of the first splice
(rear carrier), the right upper jaw is raised enough to disengage the Micro
switch controlling the heat cycle.
o A push button switch located in the
front panel of the control box is depressed. This makes the air cylinder move
the electrode holder to the front position, and lights a green light
(integrated in the switch)
o The upper right jaw is now again
lowered to complete the second splice.
Note that the
holder stays in the front position (and the green light stays on)
until the
solder starts feeding for the next splice, at which time the holder moves automatically to the rear
position and the green light goes out.
2. Control of Tungsten Carbide
Electrode Alignment.
To ensure a
long electrode life it is imperative that the two electrodes rest against the carrier to be spliced with
an even pressure. In other words: that the two active electrode surfaces are
parallel.
An un-even
pressure on the carrier will result in local overheating, causing
the carrier
metal to melt and become stuck to the electrode surface.
While it
sometimes is possible to clean the electrode surface, using emery cloth, in many cases
particles will be pulled out of the electrode surface,
with pitting
as a result. A pitted electrode surface will, at best, cause a thickness increase in the splice, but in most
cases it will require replacing the
electrode.
We have found
that by placing a piece of semi-conducting paper between the two electrodes and measuring the current
flowing across the paper, we can
determine
when the two surfaces are parallel, since this will coincide with
Issue No.8
(con'd):
-
2 -
a maximum in
the current measured on a 0 to 10 milliampere meter.
As a service
to our customers we have made up a kit consisting of a 0 to 10 mA instrument, a
1 kOhm resistor and a 9 volt battery, plus the requires leads for connection to
the electrodes.
The procedure
for using this kit is as follows:
o The power lead to the lower
electrode holder is disconnected
from the connecting block and one of the crocodile clips (from the
kit) is attached
to it. The
other crocodile clip may be fastened to the top holder.
o The other ends of these leads are
plugged into the box containing the instrument.
o To check the connections, close the
right jaw on the splicer. You should now
read 9 mA.
o Place a piece of carbon black paper
between the two electrodes. The reading should
now be somewhat below 2 mA.
o While watching the instrument, try
adjusting the position of the upper electrode
by tightening
or loosening the two 3 mm nuts on the angle just behind the electrode holder. When the maximum
reading is obtained, the two electrode surfaces
are parallel.
It is
advisable to make this control every time a new electrode is installed, and before
an attempt is
made to make a splice, since this first splice may ruin the electrode
(by making it
become pitted).
3. Wide Strip Splicer.
The new
splicer mentioned under (2) in Issue No. 6 suitable for the splicing of wide
strips in front of stamping presses has been improved further.
It is now
possible to splice strips ranging in width from 6 mm to 52 mm using the
standard machine.
To change
from one strip width to another only involves loosening two screws in each of
the two front guide rails, pushing these rails up against the new strip, and
fastening them again.
The soldering
process stops automatically when the far edge of the strip is reached, and in
most cases the other parameters may be left unchanged, so that one unit may
serve a
number of presses operating with different strip widths.
A VIDEO
cassette, which shows the operation of the splicer, is available.